【摘要】:目的探討一種最新的光學(xué)低相干反射測(cè)量?jī)xLenstar LS 900對(duì)于眼球生物測(cè)量的準(zhǔn)確性,將其測(cè)量值與光學(xué)相干生物測(cè)量?jī)xIOLMaster的測(cè)量結(jié)果進(jìn)行比較。方法對(duì)照觀察分別由Lenstar LS 900和IOLMaster測(cè)量的正常人群眼球的眼軸長(zhǎng)度(axial length,AL)、前房深度(anterior chamber depth,ACD)、角膜子午線上下的曲率半徑(CR1、CR2)和白對(duì)白角膜直徑(white-to-white distance,WTW)等眼球生物學(xué)參數(shù)。結(jié)果本研究納入了206個(gè)健康的成年個(gè)體(206眼),年齡(26.9±15.6)歲。兩種測(cè)量?jī)x對(duì)于AL、ACD、CR1、CR2和WTW等測(cè)量值的相關(guān)程度很高。Lenstar LS 900與IOLMaster的測(cè)量結(jié)果比較,AL差值(0.014±0.090)mm(P=0.025),CR1差值(0.006±0.030)mm(P=0.005),CR2差值(0.020±0.040)mm(P0.001),均略長(zhǎng);WTW較短差值(-0.110±0.400)mm(P0.001),ACD的差異無(wú)統(tǒng)計(jì)學(xué)意義(P=0.554)。除WTW一致性界限為-0.89~0.67 mm外,兩種測(cè)量?jī)x的測(cè)量數(shù)據(jù)具有良好的一致性。結(jié)論與IOLMaster相比較,Lenstar LS 900可提供同樣準(zhǔn)確、可靠的眼球生物測(cè)量數(shù)據(jù)(除WTW外),并且可提供中央角膜厚度和晶狀體厚度數(shù)據(jù),因此Lenstar LS 900在屈光手術(shù)和白內(nèi)障手術(shù)領(lǐng)域有很好的應(yīng)用前景,能滿足未來(lái)高質(zhì)量白內(nèi)障手術(shù)的要求。
[Abstract]:Objective to investigate the accuracy of a new optical low coherence reflection measuring instrument (Lenstar LS 900) for eyeball bioassay, and to compare its measured value with that of optical coherent biometric instrument (IOLMaster). Methods the axial length (axial length,AL), anterior chamber depth (anterior chamber depth,ACD) and curvature radius (CR1,) of the upper and lower meridians of the cornea were measured by Lenstar LS 900 and IOLMaster, respectively. The biological parameters of eyeball, such as CR2) and white to white corneal diameter (white-to-white distance,WTW). Results 206 healthy adult individuals (206 eyes), aged (26.9 鹵15.6) years, were included in this study. The correlation between the two measuring instruments for AL,ACD,CR1,CR2 and WTW is very high. Compared with IOLMaster, AL difference (0.014 鹵0.090) mm (P 鈮,
本文編號(hào):2484106
[Abstract]:Objective to investigate the accuracy of a new optical low coherence reflection measuring instrument (Lenstar LS 900) for eyeball bioassay, and to compare its measured value with that of optical coherent biometric instrument (IOLMaster). Methods the axial length (axial length,AL), anterior chamber depth (anterior chamber depth,ACD) and curvature radius (CR1,) of the upper and lower meridians of the cornea were measured by Lenstar LS 900 and IOLMaster, respectively. The biological parameters of eyeball, such as CR2) and white to white corneal diameter (white-to-white distance,WTW). Results 206 healthy adult individuals (206 eyes), aged (26.9 鹵15.6) years, were included in this study. The correlation between the two measuring instruments for AL,ACD,CR1,CR2 and WTW is very high. Compared with IOLMaster, AL difference (0.014 鹵0.090) mm (P 鈮,
本文編號(hào):2484106
本文鏈接:http://sikaile.net/yixuelunwen/yank/2484106.html
最近更新
教材專著