自含式壓力蒸汽滅菌生物指示劑D值測定的影響因素
發(fā)布時間:2018-06-13 14:26
本文選題:自含式生物指示劑 + 嗜熱脂肪芽孢桿菌。 參考:《中國消毒學雜志》2017年07期
【摘要】:目的 研究自含式壓力蒸汽滅菌生物指示劑D值測定的影響因素,提高D值測定的準確性。方法 使用生物指示劑抗力測定儀,對不同生產(chǎn)企業(yè)的自含式壓力蒸汽滅菌生物指示劑進行D值測定,比較透氣孔、培養(yǎng)基、放置方式、載樣器材對D值測定的影響。結果 透氣孔、培養(yǎng)基、放置方式、載樣器材對D值的最大影響率分別為6.9%、17.3%、29.9%、38.2%。結論 放置方式和載樣器材是影響自含式壓力蒸汽滅菌生物指示劑D值測定的主要因素,建議用帶孔支架以水平方式進行測定。
[Abstract]:Objective to study the factors influencing the determination of D value of self contained autoclaving biological indicator and to improve the accuracy of D value determination. Methods using biological indicator resistance tester, the D value of self-contained autoclaving biological indicator of autogenous steam sterilization in different production enterprises was determined, and the effects of air permeability, culture medium, placement method and sample loading equipment on D value were compared. Results the maximum influence rates of air permeation hole, culture medium, placement method and sample loading equipment on D value were 6.9% and 17.3% and 29.9% and 38.2% respectively. Conclusion the method of placing and loading equipment are the main factors that influence the determination of D value of self-contained autoclaving biological indicator. It is suggested that the porous scaffold should be used to measure the D value of the biological indicator.
【作者單位】: 四川省食品藥品檢驗檢測院;四川科倫藥業(yè)股份有限公司;
【分類號】:R187
【相似文獻】
相關會議論文 前4條
1 詹淼;馬桂玲;;快速閱讀生物指示劑監(jiān)測滅菌效果在植入物中的應用觀察[A];全國第四屆醫(yī)院消毒供應中心護理學術交流暨專題講座會議論文匯編[C];2008年
2 詹淼;張紅玲;王曉東;;快速閱讀生物指示劑監(jiān)測滅菌效果在植入物中的應用觀察[A];消毒供應中心(室)建設與管理高級研修班暨學術會論文集(1)[C];2008年
3 周莉;秦紅梅;;自動閱讀器監(jiān)測生物指示劑的使用及結果分析[A];全國口腔科護理學術交流暨專題講座會議論文匯編[C];2005年
4 褚金萍;;1292快速生物監(jiān)測指示劑的應用體會[A];中華護理學會第5屆全國消毒供應中心發(fā)展論壇論文匯編[C];2009年
相關碩士學位論文 前1條
1 王新建;最終滅菌藥品無菌保證體系和參數(shù)放行的關鍵控制因素探討[D];山東大學;2006年