天堂国产午夜亚洲专区-少妇人妻综合久久蜜臀-国产成人户外露出视频在线-国产91传媒一区二区三区

先進核醫(yī)學影像探測器的位置和能量性能優(yōu)化研究

發(fā)布時間:2018-01-25 17:11

  本文關鍵詞: 閃爍晶體探測器 CZT DOI 雙端讀出 連續(xù)晶體 出處:《清華大學》2016年博士論文 論文類型:學位論文


【摘要】:以正電子發(fā)射斷層成像(Positron Emission Tomography,PET)和單光子發(fā)射計算機斷層成像(Single Photon Emission Computed Tomography,SPECT)為代表的核醫(yī)學影像學是現(xiàn)代醫(yī)學影像診斷的主要手段之一。伽馬光子探測器是PET和SPECT成像系統(tǒng)的核心組成部分,優(yōu)化探測器性能對提升成像系統(tǒng)綜合性能和臨床診斷價值具有重要意義。本論文旨在結合PET和SPECT影像系統(tǒng)實際臨床應用,研究現(xiàn)代核醫(yī)學影像探測器對伽馬事件的位置和能量測量性能優(yōu)化問題。能量分辨率對雙核素SPECT成像具有重要價值。針對高能量分辨率CZT探測器在SPECT中的應用,重點研究CZT不完全電荷收集過程引入的低能尾部光子計數(shù)效應及其校正方法;谏虡I(yè)CZT SPECT系統(tǒng),提出CZT的尾部光子投影計數(shù)模型,開發(fā)了99mTc/123I雙核素成像的散射和串擾校正方法,有效提高了圖像定量精度。心肌模型實驗99mTc圖像的非透壁和透壁缺陷對比度由0.39和0.62分別提高到0.51和0.73;病人實驗99mTc圖像的心肌血池對比度由7.0提高到了63.6。測量作用深度(Depth of Interaction,DOI)信息是當前PET探測器技術的重點發(fā)展方向。本論文基于雙端讀出技術,研究了陣列式探測器的三維位置定位和能量性能優(yōu)化方法。重點研究了不同晶體表面處理方式和DOI定位算法下探測器的DOI定位性能,提出了定量描述探測器能量性能的指標。最優(yōu)的晶體表面處理方式為研磨晶體表面+鏡面反射膜(能夠獲得優(yōu)于2 mm的DOI分辨率和均衡的能量性能);線性算法和對數(shù)算法均能有效提取DOI信息并具有可類比的DOI性能。相比陣列式探測器,基于連續(xù)晶體的全三維位置定位技術是核醫(yī)學影像探測器的近年研究熱點之一。本論文研究了基于連續(xù)LYSO晶體和SiPM的全三維探測器技術。分別研究基于統(tǒng)計(Statistical-Based Positioning,SBP)和基于人工神經(jīng)網(wǎng)絡(Artificial Neural Network,ANN)的三維位置定位算法,實現(xiàn)了x/y和DOI方向分別約為2.0 mm和4.0 mm的定位分辨率。針對探測器讀出通道數(shù)過多問題,提出基于Anger位置加權和基于主成分分析(Principal Component Analysis,PCA)的兩種通道數(shù)壓縮策略;利用SBP,x/y和DOI方向的定位分辨率分別約為2.6 mm和5.0 mm,與不壓縮情況的定位能力可類比,從而提高了連續(xù)晶體探測器的工程應用價值。課題綜合研究了核醫(yī)學影像探測器的位置和能量性能優(yōu)化方法,對于雙核素SPECT成像定量精度提升以及PET和SPECT的空間分辨率提升均具有重要意義。
[Abstract]:Positron Emission Tomography was obtained by positron emission tomography. Pet) and single Photon Emission Computed Tomography. Nuclear medical imaging represented by SPECT is one of the main methods of modern medical imaging diagnosis. Gamma photon detector is the core component of PET and SPECT imaging system. The optimization of detector performance is of great significance to improve the comprehensive performance and clinical diagnostic value of imaging system. This paper aims to combine the practical clinical application of PET and SPECT imaging systems. This paper studies the optimization of the position and energy measurement performance of modern nuclear medical image detectors for gamma events. The energy resolution is of great value for binuclear SPECT imaging. Application in SPECT. The low energy tail photon counting effect introduced by CZT incomplete charge collection process and its correction method are studied. Based on the commercial CZT SPECT system, the tail photon projection counting model of CZT is proposed. The scattering and crosstalk correction methods for 99mTc / 123I binuclear imaging have been developed. The contrast of non-permeable and transmural defects in 99mTc images was improved from 0.39 and 0.62 to 0.51 and 0.73 respectively. The contrast of myocardial blood pool of 99mTc images was increased from 7.0 to 63.6. The depth of action was measured. DOI) information is the focus of the current PET detector technology. This paper based on dual-terminal readout technology. The three-dimensional position location and energy performance optimization of array detector are studied, and the DOI location performance of detector under different crystal surface treatment and DOI positioning algorithm is studied. The index of quantitatively describing the energy performance of the detector is presented. The best method of crystal surface treatment is the mirror reflection film on the ground crystal surface (the DOI resolution is better than 2 mm and the energy performance is equalized). Both the linear algorithm and the logarithmic algorithm can extract DOI information effectively and have comparable DOI performance. Full-3D position location technology based on continuous crystal is one of the hot research topics of nuclear medical image detector in recent years. In this paper, the full-3D detector technology based on continuous LYSO crystal and SiPM is studied respectively. Based on statistics (. Statistical-Based Positioning. SBP) and 3D position location algorithm based on artificial Neural Network (Ann). The location resolution of x / y and DOI is about 2.0 mm and 4.0 mm respectively. Two channel number compression strategies based on Anger position weighting and Principal component Analysis (PCA) are proposed. The location resolution in the direction of SBP x / y and DOI is about 2.6 mm and 5.0 mm, respectively, which can be compared with the localization ability in the case of no compression. Therefore, the engineering application value of continuous crystal detector is improved. The optimization method of position and energy performance of nuclear medical image detector is studied synthetically. It is of great significance to enhance the quantitative accuracy of binuclear SPECT imaging and the spatial resolution of PET and SPECT.
【學位授予單位】:清華大學
【學位級別】:博士
【學位授予年份】:2016
【分類號】:R817

【相似文獻】

相關期刊論文 前10條

1 聶聰;;多層CT探測器的發(fā)展[J];中國醫(yī)學裝備;2008年07期

2 徐慶豐;周莉;李濤;彭光;傅玉川;柏森;;重力對二維探測器陣列驗證靜態(tài)調(diào)強計劃的影響[J];華西醫(yī)學;2010年12期

3 楊昆;李真;楊永鑫;王文理;任秋實;;小動物正電子發(fā)射斷層成像儀探測器發(fā)展[J];中國生物醫(yī)學工程學報;2014年02期

4 牟文斌;16層CT技術進展[J];中國醫(yī)療器械信息;2003年05期

5 郝濤;羅宏;;CT探測器的技術特點和發(fā)展趨勢[J];中國醫(yī)療設備;2008年01期

6 核臨界循環(huán)過程中輻射安全研究項目組;曲延濤;;基于Si探測器的內(nèi)照射活體測量系統(tǒng)計算機模擬[J];中國原子能科學研究院年報;2010年00期

7 J.柯斯克能 ,藺春濤,高冬;線陣探測器(LDA)的現(xiàn)狀及發(fā)展趨勢[J];CT理論與應用研究;2002年03期

8 何養(yǎng)元;;計算機斷層技術(三) 計算機斷層術的原理(2)[J];醫(yī)療器械;1981年03期

9 柳澄,秦維昌;多層螺旋CT(一)[J];醫(yī)學影像學雜志;2000年03期

10 冀敏,滑炎卿;多層螺旋CT技術的現(xiàn)狀與發(fā)展[J];中國醫(yī)學物理學雜志;2003年01期

相關會議論文 前10條

1 鄒鴻;陳鴻飛;鄒積清;施偉紅;張錄;寧寶俊;田大宇;;硅條帶探測器陣列在空間粒子探測中的應用[A];中國空間科學學會空間探測專業(yè)委員會第十八次學術會議論文集(下冊)[C];2005年

2 李齊林;鄧小武;陳立新;黃曉延;;2-D半導體探測器陣列方向性響應的研究[A];中華醫(yī)學會放射腫瘤治療學分會六屆二次暨中國抗癌協(xié)會腫瘤放療專業(yè)委員會二屆二次學術會議論文集[C];2009年

3 馬攀;戴建榮;Jie She;;探測器陣列大野測量時散射體積缺損對測量準確性的影響[A];2007第六屆全國放射腫瘤學學術年會論文集[C];2007年

4 張行坤;;數(shù)字化X線設備CR和DR的成像原理[A];中華醫(yī)學會醫(yī)學工程學分會第八次學術年會暨《醫(yī)療設備信息》創(chuàng)刊20周年慶祝會論文集[C];2006年

5 吳廣國;黃勇;賈彬;曹學蕾;孟祥承;王煥玉;李秀芝;梁琨;楊茹;韓德俊;;硅漂移探測器的制作工藝及特性研究[A];第十四屆全國核電子學與核探測技術學術年會論文集(上冊)[C];2008年

6 吳廣國;黃勇;賈彬;曹學蕾;孟祥承;王煥玉;李秀芝;梁琨;楊茹;韓德俊;;硅漂移探測器的制作工藝及特性研究[A];第十四屆全國核電子學與核探測技術學術年會論文集(1)[C];2008年

7 徐子森;;多層螺旋CT技術新進展[A];山東醫(yī)學會醫(yī)療器械專業(yè)委員會第八次學術年會論文集[C];2000年

8 L.Bagby;M.Johnson;R.Lipton;顧維新;;硅條探測器及其讀出芯片SVX飽和的研究[A];第8屆全國核電子學與核探測技術學術年會論文集(一)[C];1996年

9 向思樺;陳四海;潘峰;黃磊;賴建軍;柯才軍;易新建;;柔性仿生復眼成像系統(tǒng)探測機理模擬研究[A];2004全國圖像傳感器技術學術交流會議論文集[C];2004年

10 常勁帆;王錚;李秋菊;張研;;LHAASO地面粒子探測器陣列前端電子學初步設計[A];第十五屆全國核電子學與核探測技術學術年會論文集[C];2010年

相關重要報紙文章 前1條

1 實習生 姜靖;探索在量子領域前沿[N];科技日報;2007年

相關博士學位論文 前10條

1 石銘;延伸波長InGaAs探測器的表面與界面研究[D];中國科學院研究生院(上海技術物理研究所);2015年

2 黃星;短波紅外InGaAs探測器輻照特性研究[D];中國科學院研究生院(上海技術物理研究所);2015年

3 陳泉佑;利用大亞灣中微子探測器進行惰性中微子實驗的可行性研究[D];山東大學;2016年

4 馬躍東;共面柵碲鋅鎘探測器的性能優(yōu)化與陣列探測器研究[D];重慶大學;2016年

5 夏鑫;AMS-02實驗中原初宇宙線~3He/~4He通量比的測量[D];山東大學;2017年

6 李梁;長波/甚長波紅外量子級聯(lián)探測器相關物理研究[D];華東師范大學;2017年

7 范鵬;先進核醫(yī)學影像探測器的位置和能量性能優(yōu)化研究[D];清華大學;2016年

8 王凡;仿視網(wǎng)膜分布探測器設計及成像特性研究[D];北京理工大學;2015年

9 楊賀潤;Micromegas探測器的性能研究[D];蘭州大學;2010年

10 楊波;可見拓展的短波紅外InGaAs探測器研究[D];中國科學院研究生院(上海技術物理研究所);2014年

相關碩士學位論文 前10條

1 谷野;基于四象限探測器的跟蹤與通信復用技術研究[D];長春理工大學;2016年

2 劉祥;基于SiPM的數(shù)字化PET探測器設計、實現(xiàn)與評估[D];華中科技大學;2015年

3 王鑫;工業(yè)CT探測器光電流信號放大及轉(zhuǎn)換特性研究[D];重慶大學;2011年

4 裴常進;錦屏深地核天體物理實驗BGO探測器的研制[D];鄭州大學;2017年

5 齊國濤;Si-PIN X射線探測器性能研究[D];河北師范大學;2012年

6 劉海;LiTaO_3探測器的二維結構分析[D];長春理工大學;2007年

7 范生強;BST鐵電薄膜紅外探測器陣列的制備工藝及其電學性能[D];湖北大學;2011年

8 郝博濤;大氣激光通信系統(tǒng)探測器陣列分集接收技術研究[D];長春理工大學;2013年

9 倉黎黎;基于384×288探測器的非制冷紅外圖像處理技術研究[D];南京理工大學;2009年

10 丁勇;軌道振動和探測器傾斜對錐束CT的影響[D];東北大學;2008年

,

本文編號:1463275

資料下載
論文發(fā)表

本文鏈接:http://sikaile.net/yixuelunwen/yundongyixue/1463275.html


Copyright(c)文論論文網(wǎng)All Rights Reserved | 網(wǎng)站地圖 |

版權申明:資料由用戶cfc28***提供,本站僅收錄摘要或目錄,作者需要刪除請E-mail郵箱bigeng88@qq.com