SUN醫(yī)用工作站NVRAM信息丟失的處理方法
發(fā)布時(shí)間:2018-03-03 08:54
本文選題:NVRAM 切入點(diǎn):SUN 出處:《臨床放射學(xué)雜志》2014年04期 論文類(lèi)型:期刊論文
【摘要】:正故障現(xiàn)象GE 8排螺旋CT后處理工作站(主機(jī)型號(hào)為ultra60,軟件版本為AW 4.0)開(kāi)機(jī)后,硬件自檢完成后系統(tǒng)終止運(yùn)行,操作系統(tǒng)不能啟動(dòng)。故障分析仔細(xì)閱讀監(jiān)視器上的信息,發(fā)現(xiàn)在屏幕下方顯示的信息異常:Ethernet address 00:00:00:00:00:00,Host ID 00010000.The IDPROM contents are invalid.強(qiáng)制關(guān)機(jī)后再開(kāi)機(jī),現(xiàn)象依舊。反復(fù)開(kāi)機(jī)重啟發(fā)現(xiàn)Eth-
[Abstract]:Positive fault phenomenon GE 8 slice spiral CT post-processing workstation (host type ultra60, software version AW 4.0) is turned on, the system stops running after the completion of the hardware self-test, the operating system cannot start, and the fault analysis carefully reads the information on the monitor. Find the abnormal information at the bottom of the screen:: Ethernet address 00: 00: 00: 00: 00: 00: 00: 00: 00: 00: 00: 00: 00: 00: 00: 00: 00: 00: 00: 00: 00: 00: 00: 00: 00: 00: 00: 00: 00: 00: 00: 00: 00: 00: 00: 00: 00: 00:
【作者單位】: 寧夏醫(yī)科大學(xué)總醫(yī)院放射科;寧夏醫(yī)科大學(xué)心腦血管醫(yī)院放射科;
【分類(lèi)號(hào)】:R197.39
【相似文獻(xiàn)】
相關(guān)期刊論文 前10條
1 趙正元;西門(mén)子P_8型磁共振NVRAM數(shù)據(jù)錯(cuò)亂的處理方法[J];云南大學(xué)學(xué)報(bào)(自然科學(xué)版);2000年S2期
2 鄭蘇芬;MLP—190激光相機(jī)維修實(shí)例[J];醫(yī)療設(shè)備信息;2003年05期
3 黃煌鏡,劉曼芳,黃 剛;M952型激光相機(jī)常見(jiàn)一例故障的排除[J];中國(guó)醫(yī)療器械雜志;2002年04期
4 黃煌鏡,劉曼芳;LC/LP DSA減影系統(tǒng)常見(jiàn)故障的檢修[J];醫(yī)療設(shè)備信息;2004年05期
5 金匡智;柯達(dá)1120激光相機(jī)故障分析[J];醫(yī)療設(shè)備信息;1999年05期
6 陸蓉,單紅旭,任恕;多導(dǎo)高精度大容量數(shù)據(jù)記錄儀的研制[J];醫(yī)療衛(wèi)生裝備;1998年05期
7 ;[J];;年期
8 ;[J];;年期
9 ;[J];;年期
10 ;[J];;年期
,本文編號(hào):1560359
本文鏈接:http://sikaile.net/yixuelunwen/swyx/1560359.html
最近更新
教材專(zhuān)著