天堂国产午夜亚洲专区-少妇人妻综合久久蜜臀-国产成人户外露出视频在线-国产91传媒一区二区三区

電子式和機械式兩種吸氣峰流速測量方法比較

發(fā)布時間:2018-08-10 20:38
【摘要】:目的吸入療法在呼吸疾病的診療中廣泛應用,其中干粉吸入器(DPI)是最常用的吸入裝置,因吸氣驅動,須達到足夠的吸氣流量。本課題旨在利用In-Check DIAL吸氣峰流量儀模擬各種DPI的內部阻力,比較電子式和機械式兩種測量吸氣峰流量(PIF)的方法,觀察不同氣道阻塞程度患者的吸氣流量特征,為臨床選擇PIF測量方法提供客觀依據。方法采用自身對照試驗的方法,以不同嚴重程度的氣道阻塞患者作為研究對象,均采用電子式和機械式兩種方法同步測量PIF。以串聯(lián)的形式將電子式測量儀(PowerLab ML870的1000L流量頭)和機械式測量儀(In-Check DIAL吸氣峰流量儀)相連接,指導受試者先呼氣至殘氣位,然后口含咬口器,再從連接好的測定裝置中進行用力快速吸氣,并吸氣至肺總量位,記錄電子式測量儀和機械式測量儀的同步測量值。一共測量3種不同阻力狀態(tài)下的PIF:無阻力狀態(tài)(Free)、模擬準納器阻力狀態(tài)(Accuhaler)和模擬都保阻力狀態(tài)(Turbuhaler)。每一種阻力狀態(tài)下至少測定3次,重復性要求≤10%,分別取電子式和機械式測量的最佳值用于計算。結果共納入30例氣道阻塞患者,依肺功能障礙的程度,分為輕度、中度和重度阻塞3個組,年齡(67.7±7.6)歲。在Free和Accuhaler的阻力狀態(tài)下,分別有12例(40%)和1例(3.3%)因超過機械式流量儀的量程,不能獲得測量值,其余均可獲得測量值用于統(tǒng)計。(1)輕度阻塞組,機械式和電子式測量法在Free、Accuhaler和Turbuhaler阻力下的PIF(PIF-F、PIF-A、PIF-T)分別為(80.00±34.52)L/min、(76.86±32.85)L/min(P=0.051);(73.00±31.29)L/min、(71.63±31.53)L/min(P=0.126);(61.50±23.10)L/min、(59.10±22.42)L/min(P=0.075),3種阻力下兩種方法間均無統(tǒng)計學差異。(2)中度阻塞組,機械式和電子式測量法的PIF-F、PIF-A、PIF-T分別為(86.00±26.32)L/min、(82.96±29.66)L/min(P=0.027);(79.50±24.66)L/min、(76.26±24.65)L/min(P=0.028);(67.00±18.59)L/min、(62.89±17.92)L/min(P=0.008),3種阻力下兩種方法間均有統(tǒng)計學差異。(3)重度阻塞組,機械式和電子式測量法的PIF-F、PIF-A、PIF-T分別為(85.00±20.25)L/min、(84.22±22.11)L/min(P=0.599);(75.56±24.81)L/min、(74.53±25.67)L/min(P=0.374);(68.50±21.09)L/min、(65.40±19.97)L/min(P=0.016),PIF-F和PIF-A兩種方法間無統(tǒng)計學差異,PIF-T兩種方法間有統(tǒng)計學差異。結論盡管兩種測量方法存在一定的差異,但差異小于患者的自身變異,因此兩種測量方法的結果均可用于指導臨床。通過比較,電子式測量方法量程大,精度高,可反復校正調零,準確度更佳,并可記錄下呼吸波形,可計算吸氣時間以及吸氣容積等,適于深入開展科研。但電子式測量儀價格昂貴,結構復雜,體積較大,而機械式吸氣峰流量儀價格便宜、輕便易帶,操作簡單。如需外出調查或僅為患者在家自測PIF,則建議選擇機械式測量方法。
[Abstract]:Objective inhalation therapy is widely used in the diagnosis and treatment of respiratory diseases, in which dry powder inhaler (DPI) is the most commonly used inhaling device. The purpose of this study was to simulate the internal resistance of various DPI by In-Check DIAL peak inspiratory flow meter, compare the two methods of measuring the inspiratory peak flow (PIF) with electronic and mechanical methods, and observe the inspiratory flow characteristics of patients with different degree of airway obstruction. To provide objective basis for clinical selection of PIF measurement methods. Methods the PIFs of patients with different severity of airway obstruction were measured by both electronic and mechanical methods. The electronic meter (1000L flow head of PowerLab ML870) and the mechanical meter (In-Check DIAL peak inspiratory flow meter) were connected in series to guide the subjects to exhale to the residual air level first, and then to have the mouthbite device in their mouth. Then inhale quickly and forcefully from the connected measuring device and inhale to the total lung position and record the synchronous measurements of electronic and mechanical measuring instruments. A total of 3 PIFs were measured under different resistance states: no drag state (Free), simulation quasi-sensor resistance state (Accuhaler) and simulation both resistive state (Turbuhaler). At least 3 times in each resistance state, the repeatability requirement is less than 10. The optimum values of electronic and mechanical measurements are taken for calculation respectively. Results A total of 30 patients with airway obstruction were divided into 3 groups according to the degree of pulmonary dysfunction: mild, moderate and severe obstruction, aged (67.7 鹵7.6) years. Under the resistance state of Free and Accuhaler, 12 cases (40%) and 1 case (3.3%) could not obtain the measured value because they exceeded the measuring range of the mechanical flow meter. The PIF (PIF-FU) PIF-An PIF-T of mechanical and electronic methods were (80.00 鹵34.52) L / min, (76.86 鹵32.85) L/min (P0.051); (73.00 鹵31.29) L / min, (71.63 鹵31.53) L/min (P0.126); (61.50 鹵23.10) L / min, (59.10 鹵22.42) L/min (P0.075) and (59.10 鹵22.42) L/min (P0.075) L / min, respectively. The PIF-FU PIF-An PIF-T of mechanical and electronic methods were (86.00 鹵26.32) L / min, (82.96 鹵29.66) L/min (P0.027); (79.50 鹵24.66) L / min, (76.26 鹵24.65) L/min (P0.028); (67.00 鹵18.59) L / min, (62.89 鹵17.92) L/min (P0. 008), respectively. The PIF-FU PIF-An PIF-T of mechanical and electronic measurement methods were (85.00 鹵20.25) L / min, (84.22 鹵22.11) L/min (P0. 599); (75.56 鹵24.81) L / min, (74.53 鹵25.67) L/min (P0.374); (68.50 鹵21.09) L / min, (65.40 鹵19.97) L/min (P0.016) PIF-F and PIF-A respectively. Conclusion although there is a certain difference between the two methods, the difference is less than that of the patient's own variation, so the results of the two methods can be used to guide the clinical practice. By comparison, the electronic measurement method has the advantages of large measuring range, high precision, repeatable correction and zero adjustment, better accuracy, and the recording of respiratory waveform, calculation of inspiratory time and volume, etc., which is suitable for further scientific research. However, the electronic measuring instrument is expensive, complex in structure and large in volume, while the mechanical inspiratory peak flow meter is cheap, light and easy to take, and easy to operate. If you need to go out to investigate or only test PIFs at home, it is recommended to choose mechanical measurement methods.
【作者單位】: 廣州醫(yī)科大學;廣州醫(yī)科大學附屬第一醫(yī)院廣州呼吸疾病研究所(呼吸疾病國家重點實驗室呼吸疾病國家臨床研究中心);
【分類號】:R56

【相似文獻】

相關期刊論文 前1條

1 李慧文;電灼下鼻甲治療慢性單純鼻炎鼻塞[J];現(xiàn)代診斷與治療;1997年03期

,

本文編號:2176137

資料下載
論文發(fā)表

本文鏈接:http://sikaile.net/yixuelunwen/huxijib/2176137.html


Copyright(c)文論論文網All Rights Reserved | 網站地圖 |

版權申明:資料由用戶61dff***提供,本站僅收錄摘要或目錄,作者需要刪除請E-mail郵箱bigeng88@qq.com