高效液相色譜串聯(lián)質(zhì)譜法研究環(huán)磷酰胺線性微透析探針體內(nèi)外校正影響因素
發(fā)布時間:2018-05-30 17:45
本文選題:微透析 + 環(huán)磷酰胺; 參考:《國際藥學(xué)研究雜志》2015年03期
【摘要】:目的建立環(huán)磷酰胺線性探針體內(nèi)外回收率最優(yōu)校正方法。方法以復(fù)方氯化鈉注射液為灌注液,反向透析校正方法中的增量法與減量法校正線性探針,考察不同灌注流速對環(huán)磷酰胺線性探針回收率的影響。結(jié)果灌注流速為1.0μl/min時,環(huán)磷酰胺回收率大幅提高,增量法與減量法回收率分別由(24.11±1.75)%、(26.19±0.76)%提升至(50.93±0.91)%、(53.31±1.23)%,體內(nèi)回收率為(26.19±0.76)%。結(jié)論環(huán)磷酰胺適宜進行微透析技術(shù)的取樣,本實驗校正方法簡便有效。
[Abstract]:Objective to establish the optimal correction method for the recovery rate of cyclophosphamide linear probe in vivo and in vitro. Method using Compound Sodium Chloride Injection as the perfusion liquid, the increment method and the reduction method in the reverse dialysis correction method to correct linear probe, and investigate the effect of different perfusion flow velocity on the recovery rate of cyclophosphamide linear probe. The recovery rate of phosphamide was greatly improved, the recovery rate of the incremental method and the reduction method were (24.11 + 1.75)%, (26.19 + 0.76)%, (50.93 + 0.91)%, (53.31 + 1.23)% and (26.19 + 0.76)%, and the recovery rate in the body was (26.19 + 0.76)%. Conclusion cyclophosphamide is suitable for microdialysis sampling. The correction method of this experiment is simple and effective.