多氣源混輸管道在線氣相色譜儀安裝位置的確定方法
本文選題:普光氣田 + 普光首站。 參考:《天然氣工業(yè)》2017年10期
【摘要】:采用在線氣相色譜儀實時監(jiān)測管道中天然氣組分的變化是目前大型輸氣場站保證天然氣計量準(zhǔn)確性的關(guān)鍵工藝,而在線氣相色譜儀抽取樣品的代表性則是保證分析檢測結(jié)果準(zhǔn)確可靠的關(guān)鍵因素,對于多氣源匯合的輸氣場站,不同氣源天然氣的混合程度直接關(guān)系到在線氣相色譜儀所取樣品的代表性。為此,以中石化川氣東送工程的普光首站為例,按照該站的工況條件,采用FLUENT軟件對多氣源天然氣的混合規(guī)律進(jìn)行了仿真模擬,探討了多氣源天然氣混合距離與溫度、壓力、流速等因素的關(guān)系,以及多氣源天然氣混合均勻的最近位置。試驗結(jié)果表明:(1)含兩種不同摩爾分?jǐn)?shù)值組分的天然氣在匯管不同位置處混合的均勻程度不同——距匯管50 m處天然氣各組分混合不均勻,具明顯分層現(xiàn)象,距匯管150 m處天然氣各組分的混合開始呈現(xiàn)均勻狀態(tài),但仍存在分層現(xiàn)象,距匯管280 m處天然氣各組分混合完全呈現(xiàn)出均勻狀態(tài),其CH_4摩爾分?jǐn)?shù)為98.30%,CO_2摩爾分?jǐn)?shù)為0.95%;(2)在普光首站距混合點280 m以遠(yuǎn)處的位置安裝在線氣相色譜儀取樣較為合適;(3)對于氣源體積流量或氣源的體積流量差別較大、壓力較高或溫度較低的工況條件,在線氣相色譜儀的取樣點可適當(dāng)靠近混合點。
[Abstract]:On-line gas chromatograph is the key process to ensure the accuracy of natural gas measurement in large gas transmission stations. On the other hand, the representativeness of sampling samples by on-line gas chromatograph is the key factor to ensure the accuracy and reliability of the analysis and detection results. The mixing degree of natural gas from different gas sources is directly related to the representativeness of the samples taken by on-line gas chromatograph. Therefore, taking Puguang first station of Sinopec Chuan Gas East Transmission Project as an example, according to the working conditions of the station, the mixing law of multi-gas source natural gas is simulated by fluent software, and the mixing distance, temperature and pressure of multi-gas source natural gas are discussed. The relationship between flow rate and other factors, as well as the nearest position of multiple gas source natural gas mixing uniformity. The experimental results show that: (1) the mixing uniformity of natural gas containing two different molar numerical components is different at different locations of the sink pipe, and the mixture of natural gas components is not uniform at 50 m away from the sink tube, and there is obvious stratification phenomenon. The mixture of each component of natural gas begins to show uniform state at 150 m from sink tube, but there still exists stratification phenomenon, and the mixture of each component of natural gas shows uniform state completely at 280 m from sink pipe. Its Ch _ 4 mole fraction is 98.30 and the CO _ 2 molar fraction is 0.95; (2) it is more appropriate to install an on-line gas chromatograph to sample at a distance of 280 m from the mixing point at Puguang first station; (3) for gas source volume flow or gas source volume flow, When the pressure is high or the temperature is low, the sampling point of on-line gas chromatograph can be close to the mixing point.
【作者單位】: 中國石化天然氣分公司計量研究中心;
【基金】:中國石油化工股份有限公司科研項目“中國石化天然氣能量計量先導(dǎo)試驗研究”(編號:714024)
【分類號】:O657.71;TE832.2
【相似文獻(xiàn)】
相關(guān)期刊論文 前10條
1 郭建民,傅國林,王曉寧,郝維,劉劍;如何保證在線氣相色譜儀的分析準(zhǔn)確性[J];化工自動化及儀表;2000年05期
2 劉玉華;張廣明;;一種教學(xué)用簡易氣相色譜儀的試制[J];鄭州大學(xué)學(xué)報(自然科學(xué)版);1976年01期
3 陳裕森;關(guān)于102-G氣相色譜儀高速化的探討[J];色譜;1986年Z1期
4 孟品佳;;氣相色譜儀使用中的幾個問題[J];分析儀器;1993年03期
5 ;新型氣相色譜儀和檢測方法[J];化工文摘;2003年08期
6 陳海元;王永健;;淺談氣相色譜儀使用中容易忽視的問題[J];科技咨詢導(dǎo)報;2007年14期
7 商登喜;;氣相色譜儀使用中的幾個問題[J];石油化工;1981年04期
8 周傳玢;102G型氣相色譜儀分析三十烷醇初探[J];湖南化工;1982年02期
9 陳志敏;超靈敏光電離檢測器的便攜式氣相色譜儀[J];色譜;1985年04期
10 趙國宏;呂祖芳;;氣相色譜儀簡易冷柱上進(jìn)樣裝置[J];分析儀器;1993年04期
相關(guān)會議論文 前2條
1 姜德嶺;;淺談氣相色譜儀操作中應(yīng)注意的幾個問題(摘要)[A];全國鐵路第五屆疾病控制學(xué)術(shù)研討會論文集[C];2004年
2 ;Sigma 1型氣相色譜儀及其性能評價[A];天津市色譜研究會第一屆學(xué)術(shù)報告會論文、報告(摘要)匯集[C];1980年
相關(guān)碩士學(xué)位論文 前1條
1 邱棟;惡臭硫系氣體便攜式快速氣相色譜組件的制備與研究[D];電子科技大學(xué);2014年
,本文編號:2058652
本文鏈接:http://sikaile.net/kejilunwen/shiyounenyuanlunwen/2058652.html